Skip to main content
LightingLab / UV calibration

Dymax ACCU-CAL 50V calibration in Europe

Instrument: visible curing radiometer.

For European production and quality-assurance teams, calibration should reflect the actual instrument configuration and its use in the process rather than the product name alone.

Measurement-specific assessment

Why this model needs individual assessment

The exact detector configuration and the UV source used in production must be identified before selecting a reference.

Details to include with the instrument

Confirm the 50V version, relevant wavelength response from its datasheet and whether the displayed quantity is irradiance or integrated exposure.

Practical process consideration

Report the application lamp and exposure geometry so the laboratory can assess whether the calibration condition is representative.

Calibration versus conformity assessment

A calibration reports measurement results and their uncertainty under defined conditions. A pass/fail statement additionally requires a specified tolerance and decision rule, including any agreed guard band. Neither a conformity statement nor accredited coverage is assumed for this instrument.

Manufacturer instrument information

Spectral response: Confirm exact spectral response with serial number.

Typical application: Visible curing measurement; not automatically an UV calibration.

Manufacturer product information (specification source; not a LightingLab accreditation statement).

Calibration planning

For irradiance, specify wavelength or source spectrum, measuring range, geometry and required reporting units. For dose or radiant exposure, also specify exposure time and integration behaviour. Spectral mismatch between calibration source and process lamp can be material, especially for narrowband LEDs.

Accreditation and feasibility: ISO/IEC 17025 accreditation applies only to the measurement quantities, ranges and methods covered by the laboratory's current accredited scope. Availability for this exact model, sensor, channel and dose function must be confirmed before an accredited quotation. No calibration certificate or performance claim is implied by this page.

Request a quotation

Send the manufacturer, model, serial number, detector/sensor type, required wavelengths, irradiance and/or dose range, and desired turnaround time. Contact LightingLab.

Explore other UV calibration services

Accredited calibration availability is confirmed for the specific instrument configuration, measurand and measurement range against the laboratory’s current scope. No particular model is automatically covered.