Opsytec RM-12 calibration in Europe
Instrument: UV radiometer with external sensor.
For European production and quality-assurance teams, calibration should reflect the actual instrument configuration and its use in the process rather than the product name alone.
Measurement-specific assessment
Why this model needs individual assessment
Instrument and probe together form the measurement configuration; calibration of the display alone does not establish UV sensitivity.
Details to include with the instrument
Supply the probe model, sensor serial number, cable configuration and intended wavelength range.
Practical process consideration
Where process lamps differ spectrally from the calibration reference, assess the expected effect on readings.
Calibration versus conformity assessment
A calibration reports measurement results and their uncertainty under defined conditions. A pass/fail statement additionally requires a specified tolerance and decision rule, including any agreed guard band. Neither a conformity statement nor accredited coverage is assumed for this instrument.
Manufacturer instrument information
Spectral response: Sensor-dependent; check sensor identification.
Typical application: UV irradiance and detector configuration.
Manufacturer product information (specification source; not a LightingLab accreditation statement).
Calibration planning
For irradiance, specify wavelength or source spectrum, measuring range, geometry and required reporting units. For dose or radiant exposure, also specify exposure time and integration behaviour. Spectral mismatch between calibration source and process lamp can be material, especially for narrowband LEDs.
Accreditation and feasibility: ISO/IEC 17025 accreditation applies only to the measurement quantities, ranges and methods covered by the laboratory's current accredited scope. Availability for this exact model, sensor, channel and dose function must be confirmed before an accredited quotation. No calibration certificate or performance claim is implied by this page.
Request a quotation
Send the manufacturer, model, serial number, detector/sensor type, required wavelengths, irradiance and/or dose range, and desired turnaround time. Contact LightingLab.